Viruzyme® Ve is a high‑performance biological cleaner formulated with a powerful blend of five enzymes to ensure thorough decontamination of medical, surgical, dental, and endoscopic instruments. Designed to break down and remove even the most stubborn organic contaminants, it delivers exceptional cleaning results while protecting delicate instrument surfaces.
This Class I medical device is particularly effective at dissolving high levels of proteins, fats, and organic residues—including dried blood and tissue—making it an essential part of any modern infection‑control protocol. Safe for use on all surfaces, Viruzyme® Ve provides reliable cleaning performance at both ambient and elevated temperatures.
Compatible with manual cleaning, ultrasonic baths, washer/disinfectors, and Automated Endoscope Reprocessors (AER), it supports consistent, efficient cleaning across all stages of instrument reprocessing.
Key Features
-
Highly effective biological cleaner for proteinaceous and organic contaminants
-
Five‑enzyme formula: protease, amylase, lipase, cellulase, and mannanase
-
Suitable for sensitive instrument materials
-
Pleasant odour for improved user comfort
-
Excellent material compatibility
-
Non‑fixative and non‑corrosive—protects delicate coatings
-
Rapid cleaning action: 5‑minute immersion at 20–55°C
-
Cost‑effective solution for routine decontamination
Application
Manual & Ultrasonic Cleaning of Instruments and Equipment
-
Standard dosage: Add 2–10 ml of Viruzyme® Ve per 1 litre of water (0.2–1.0%).
-
For heavy contamination: Use up to 40 ml per litre (4%) when dealing with high levels of organic material or hazardous contaminants.
-
Soaking time: Immerse parts for 5–10 minutes in the diluted solution.
-
For dried‑on protein deposits or low‑temperature solutions, extend soaking up to 20 minutes.
-
-
After soaking:
-
Brush to remove gross debris (manual cleaning)
-
Rinse thoroughly with clean water
-
Or transfer directly to the next cleaning stage (washer/disinfector, ultrasonic unit, or AER)
-
Manual Cleaning Parameters
-
Water: Warm tap water
-
Time: 5–20 minutes
-
Temperature: Ambient
-
Dosage: 2–10 ml/L (0.2–1.0%)
Pair this with 'Medistel' for a two-step process to clean and disinfect your endoscopy equipment.